Abstract

Abstract Background/purpose Cone beam computed tomography (CBCT) is frequently used in dental diagnosis and treatment. Comparative studies of the effects of CBCT on implant navigation, however, are still limited. The objective of this study was to evaluate whether the computed tomography images of the four commercial brands will affect the accuracy of the new version of IRIS implant navigation system. Materials and methods In the first part, the accuracy of the IRIS implant navigation system was evaluated by a precision confirmation jig whose position is confirmed. In the second part, the IRIS implant navigation system was used in conjunction with 4 brands of CBCT scans analyzed by its effect on accuracy. Results The results showed that the mean deviation of the new version of IRIS-100 system accuracy was less than 1 mm. Among the four groups, the overall average deviation caused by CBCT images showed that the 3D eXam group had the smallest error of approximately 0.94 ± 0.12 mm and the AZ 3000 CT group had the largest error of approximately 1.34 ± 0.10 mm. Conclusion Based on the study, the accuracy of the IRIS implant navigation system will vary with the CBCT image resolution and the status of the CBCT machine.